AstellBio Continuous Flow EDS treat biologically hazardous liquid waste while it passes through the system. The thermal process is designed to deactivate microbes, viruses, spores and genetically modified material without chemical dosing. These systems are intended for predictable or consistently high wastewater flows containing minimal solids.
The systems are intended for Biosafety Level 2 (BSL2) facilities handling biological agents that present a moderate hazard and Biosafety Level 3 (BSL3) facilities working with agents that may cause serious or potentially lethal disease, particularly through inhalation. A buffer tank compensates for variations in incoming flow before the wastewater is pumped through heat exchangers, sterilized and cooled. Astell systems are made in the UK by a sterilization equipment manufacturer established in 1884.
All configurations operate at 140 °C with a 1-minute holding time and are intended for effluent with minimal solids.
Unlike batch systems, the Continuous Flow EDS does not collect each processing volume in a sterilization vessel. Effluent is treated continuously, making the system suitable for sustained flow rates and short periods of increased demand.
All configurations operate at 140 °C with a 1-minute holding time. The defined relationship between flow rate, temperature and holding time provides controlled treatment as the wastewater passes through the equipment.
Continuous-flow systems are intended for liquid effluent with minimal solids. Significant particulate content can foul the heat exchangers and block the system’s narrow pipework, so wastewater containing suspended solids should be directed to a batch EDS instead.
Incoming wastewater is preheated using heat recovered from the treated effluent. This process can recover up to 80% of the heat energy.
The buffer tank balances variations in wastewater production before the liquid is pumped through the treatment system at a controlled rate. Tank capacity increases with the selected system throughput.